Метанатор для GC-2030
Описание
- Converts CO and CO2 to Methane
- Wide linear range from 0.1 ppm to 100%
- Unique, 3D printed non-nickel catalyst for more resistance to oxygen and sulfur poisoning
- Directly replaces existing FID jets; no additional fittings or connections required
- Uses existing FIG fuel gases; no additional gas flows required
- Available as a single or multipack
| Content | P/N |
|---|---|
| 1 pack | 220-94673-30 |
| 3 packs | 220-94673-99 |
| 5 packs | 220-94673-35 |
Performance Specifications
| Minimum Detection Limit | 1 pg C/sec. |
|---|---|
| Linear Dynamic Range | > 106 |
| Accuracy | < 3% standard error |
| Max. Injection Volume | 1 ml (10:1 split) |
| Max. Temperature | 450° C |
| Response Time | Accomodates 500 Hz FID detectors |
| Detection Precision | < 1% RSD |
| Lifetime | Usage and analyte dependent. Optimal performance for 6 months before recommended replacement |
Typical Operating Parameters
| Fuel Gas Ratio |
H2: 32 ml/min. |
|---|---|
| Temperature | 400° C |
Connectivity
| Capillary Column Sizes | All columns 50 µm to 530 µm ID |
|---|---|
| Packed Column Sizes | 1/4″ or 1/8″ (with adapter) |
| FID Connections | Capillary, 1/8″, or 1/4″ connection (with adapter) |
| Sampling | Gas sampling, manual injection, autosampler, total evaporation headspace |
Adapters
| # | Description | P/N |
|---|---|---|
| 1 |
1/8″ to 1/16″ reducing union
1/4″ to 1/16″ reducing union
|
980-27455
980-27454
|
| 2 | Fused silica adapter for 0.53 mm ID capillary (pack of 5) | 980-05636 |
| 3 | SH-Rxi Guard Column; 0.53 mm ID; 5 m* | 227-36307-01 |
| 4 | Graphite ferrule for 0.53 mm ID capillary (pack of 10) | 221-32126-08 |
* Should be cut to the desired length