SAMPLE INTRODUCTION – GC Inlet Liners

Описание

The GC inlet functions as the interface between the syringe and the GC capillary column, where the sample is introduced, vaporized, mixed with carrier gas and transferred to the column. Shimadzu instruments offer several types of inlets — split, splitless, programmable temperature vaporization (PTV) and on-column.

The inlet liner prevents the sample contacting the metal walls of the injector block. Inlet liner geometry and packing materials enable the inlet liner to achieve greater heated surface area; this additional surface area can often improve sample vaporization. Conversely, choosing the wrong inlet liner geometry can significantly decrease the reproducibility and quality of analysis.

Injection Technique Sample Types Inlet Liner Geometry Function
Splitless

Trace Level Analyses/

Active Compounds

Taper
A bottom taper focuses sample onto the head of the column and minimizes sample contact with metal parts of the inlet. The addition of quartz wool to your inlet liner promotes mixing of analytes, aids the vaporization of liquid samples, and works as a trap to collect non-volatile residue in the sample (i.e. protects capillary column from «dirty» samples).
Split

General Purpose/

Concentrated Samples/

Dirty Samples

FocusLinerTM
Ensures quartz wool remains in the correct position in the liner. Excellent reproducibility results from the wiping of the sample from the syringe needle and the prevention of droplet formation. Minimizes high molecular weight discrimination.
Splitless

Trace Level Analyses/

Dirty Samples/

Wide Boiling Point Range

Taper FocusLinerTM
Bottom taper focuses sample onto the head of the column and minimizes contact with metal parts of the inlet. Excellent reproducibility results from the wiping of the sample from the syringe needle and the prevention of droplet formation. Minimizes high molecular weight discrimination.
Direct

Trace Level Analyses/

Active Compounds

Direct Taper
Direct inlet liners facilitate maximum transfer of sample by connecting directly to the GC column and inhibiting sample degradation due to hot metal components inside inlet.
Split/Splitless

General Purpose/

Concentrated Samples/

Dirty Samples (only if quartz wool is present)/

Gaseous Samples (also purge & trap, headspace)

Straight
Straight inlet liners facilitate higher split flows. Narrow bore straight inlet liners facilitate fast GC work. Small injection volumes of less than 0.5 μL are best used with a narrow bore. Narrow bore straight inlet liners improve focussing of gaseous samples (purge, trap & headspace).

Inlet Liner Deactivation
Every batch of inlet liners are tested for inertness using the EPA 8081B method. This standard method ensures that each batch of inlet liners has less than 3 % Endrin breakdown from a 1 ppm injection.

Glass Insert for Capillary Column Analysis
Glass insert is very important to prevent the column from being contaminated by sample components. Please choose an appropriate insert according to your system model, injection port and injection method, and exchange it regularly to avoid poor reproducibility and peak shape caused by the crossover of residual samples, etc.

S: Standard; O: Option

P/N Specification GC-2030
GC-2010 Plus
GC-2010
GC-2014 GC-2025 GC-17A Ver.1-3
GC-1700, GC-18A
GC-14A
GC-14B
GC-8A
Injection Port Side Split/
Splitless
Injection Method
Injection Unit → SPL-2030
SPL-2010 Plus
SPL-2010
SPL-2014 SPL-2025 SPL-17 SPL-14 SPL-G9
For
Split
227-35007-01
 
 (95mm)
S
221-41444 For Split
(95mm)
S S S
221-41444-01 For Split
(95mm)
S
221-37574-01 For Split
(99mm)
S
For
Splitless
227-35008-01
 
 (95mm)
S
221-48335-01 For Splitless
(95mm)
S S O
221-41544 For Splitless
(95mm)
O O O S
221-75192 For Splitless
(95mm, deactivated, 5 pcs)
O O O
221-32544 For Splitless
(99mm)
S
For
Split
and Splitless
221-75187 For Split and Splitless
(95mm, deactivated, with wool, 5 pcs)
O O
221-75188 For Split and Splitless
(95mm, deactivated, needle onto wool , 5 pcs)
O O
221-75189 For Split and Splitless
(95mm, deactivated, needle into wool , 5 pcs)
O O
221-75190 For Split and Splitless
(95mm, deactivated, 5 pcs)
O O
221-75191 For Split and Splitless
(95mm, deactivated, with wool, 5 pcs)
O O
221-75193 For Split and Splitless
(95mm, deactivated, with wool, 5 pcs)
S
221-75194 For Split and Splitless
(95mm, deactivated, 5 pcs)
O
221-41444-05 For Split and Splitless
(95mm, deactivated, 5 pcs)
S S S
221-41544-05 For Split and Splitless
(95mm, deactivated, 5 pcs)
O O S
221-75195 For Split and Splitless
(95mm, deactivated, with wool, 5 pcs)
S S O
Direct
Injection
Method
(WBI)
Injection Unit →
WBI-2030
WBI-2010 Plus
WBI-2010
WBI-2014 WBI-17
221-41599 Ddirect Injection Method(WBI)
(95mm)
O O S
221-48335-01 Ddirect Injection Method(WBI)
(95mm)
S
(WBI-2010)
S S O
221-75197 Ddirect Injection Method(WBI)
(95mm, deactivated, with wool *2, 5 pcs)
O O O
221-41599-05 Direct Injection Method(WBI)
(95mm, deactivated, for 0.53 mmID column, 5 pcs)
O
For
AOC-5000
Plus
221-75196 For AOC-5000 Plus
(95mm, deactivated, 5 pcs)
O O
WBC
Attachment
221-38107 *1 WBC Attachment
(139mm)
S S
For
CLH-14
Injection
Port Side
221-32998-01 For CLH-14
(123mm)
S
Detector
Side
221-33000 For CLH-14
(71mm)
S

*1 Required if the WBC Attachment is used for a packed column injection in GC-2014. Insert for GC-14 septum purge unit is P/N: 221-38151-04.
*2 Wool is 25 mm from the top of insert.

Glass Insert for OCI / PTV

S: Standard; O: Option

P/N Specification
GC-2030
GC-2010
GC-2010 Plus
GC-17
Simple On-column Injection Method  Injection Unit → OCI-2010 *1 OCI-17 *2
221-49381-01 Simple On-column Injection Method O
221-49381-02 Simple On-column Injection Method
(deactivated)
O
PTV Injection Method 221-42223  PTV Injection Method S
221-49300 PTV Injection Method
(quartz)
S
221-74830-09 PTV Injection Method
(for AOC-MEPS™. system, deactivated, with wool, 5 pcs)
O

*1 GC-2010 OCI requires an additional adapter (P/N: 221-49298-91).
*2 GC-17 OCI requires an additional adapter (P/N: 221-42222-91).

Glass Insert for Packed Column Analysis
When using a glass column, one may add a glass insert to prevent the column from accumulating non-volatile components. Please choose a proper insert according to the inter diameter of the glass column.
When using a stainless column, it is necessary to connect a glass insert to a 3.0 mmID column at the side of the injection unit. Please exchange it regularly to avoid poor reproducibility and peak shape caused by the crossover of residual sample, etc.
Note: Glass insert cannot be used in a GC-8A Series instrument when using glass columns.

S: Standard; O: Option

P/N Specification GC-2014 GC-17A
Ver.1-3
GC-14A/B
Injection Port Side 221-41484 Injection Port Side O
221-14093 Injection Port Side
(for 3.0 mmID column)
S S
221-14093-84 Injection Port Side
(for 3.0 mmID column, 5 pcs)
S S
221-14094 Injection Port Side
(for 2.6 mmID column)
S S
221-14094-84 Injection Port Side
(for 2.6 mmID column, 5 pcs)
S S

*GC for packed column analysis includes one insert for a 3.0 mmID column and one insert for a 2.6 mmID column as standard accessories.

Accessories for Glass Insert
Connecting Parts for Glass Insert

O-ring

Description Applicable GC P/N
Standard
(Viton, 10 pcs)
GC-2030
GC-2010 Plus
GC-2010
GC-2014
GC-2025
227-35005-01
Standard
(fluororubber, 5pcs)
036-11203-84
For split *
(graphite, 4 pcs)
221-48393-91
For splitless/WBI *
(graphite, 4 pcs)
221-47222-91
For OCI/PTV-2010
(fluororubber, 5pcs)
GC-2030
GC-2010
036-11201-84

Ferrule

Description Applicable GC P/N
For SPL-
17/1700/18A
(graphite)
GC-17A/1700/18A 221-46403-92
For SPL-14
(graphite)
GC-14A/B 221-75182

* For high temperatures
(injection port setting temperature: 300 — 450°C)

Wool

Description P/N
Deactivated Glass Wool (2 g)
Included in GC-2010 Plus/2010/2014/2025
221-48600
Quartz Wool (2 g)
Included in GC-17A/1700/18A/14A/B
201-47616-01

Shimalite Q Quartz Beads

Description P/N
25mL 670-10458-73

The quartz beads are typically packed in a glass insert to enhance the mixing property.
They can also be used in flow restrictor tubes.

Injector Liner GC-2010 (Plus) / GC-2030
Injector liner split injection for SPL-2010 Plus, SPL-2010

Part number Description PCS Details
961-01479-02  SPL liner IPdeact. w.gl.wool
Standard split liner, prepac/wp-content/uploads/2021/12/961-01479-02-liner.pdfked with glass wool, Phenylmethyl-deactivated
5 Details
961-01479-03  SPL liner IPdeact. w.CarboFrit
Standard split liner, prepacked with CarboFrit, Phenylmethyl-deactivated
5 Details
961-01479-05  SPL liner Base deact. w.gl.wool
Standard split liner, prepacked with glass wool, Base-deactivated
5 Details
961-01479-06  SPL liner Base deact. w.CarboFrit
Standard split liner, prepacked with CarboFrit, Base-deactivated
5 Details
961-01479-07  SPL liner Siltek deact. w/o packing
Standard split liner, without filling, Siltek-deactivated
5 Details
961-01479-09  SPL liner Siltek deact. w. CarboFrit
Standard split liner, prepacked with CarboFrit, Siltek-deactivated
5 Details

Injector liner splitless injection for SPL-2010 Plus, SPL-2010, WBI-2010 Plus, WBI-2010

Part number Description PCS Details
961-01480-02  SPLless liner IPdeact. w.gl.wool
Standard splitless liner, prepacked with glass wool, Phenylmethyl-deactivated
5 Details
961-01480-03  SPLless liner IPdeact. w.CarboFrit
Standard splitless liner, prepacked with CarboFrit, Phenylmethyl-deactivated
5 Details
961-01480-05  SPLless liner Base deact. w.gl.wool
Standard splitless liner, prepacked with glass wool, Base-deactivated
5 Details
961-01480-06  SPLless liner Base deact. w.CarboFrit
Standard splitless liner, prepacked with CarboFrit, Base-deactivated
5 Details
961-01480-08  SPLless liner Siltek deact. w.gl.wool
Standard splitless liner, prepacked with glass wool, Siltek-deactivated
5 Details
961-01480-09  SPLless liner Siltek deact. w. CarboFrit
Standard splitless liner, prepacked with CarboFrit, Siltek-deactivated
5 Details

Injector liner split injection for PTV-2010 , PTV-2030

Part number Description PCS Details
961-01481-02 PTV liner IPdeact. w.gl.wool
Standard PTV split liner, prepacked with glass wool, Phenylmethyl-deactivated
5 Details
961-01481-05 PTV liner Base deact. w.gl.wool
Standard PTV split liner, prepacked with glass wool, Base-deactivated
5 Details
961-01481-08 PTV liner Siltek deact. w.gl.wool
Standard PTV split liner, prepacked with glass wool, Siltek-deactivated
5 Details

Special Liners for Solid Phase Micro Extraction (SPME) for SPL-2010 Plus, SPL-2010, WBI-2010 Plus, WBI-2010 , SPL-2014, WBI-2014, SPL-2025

Part number Description PCS Details
961-01482-01  SPME liner IP deact w/o packing
Special SPME liner, without filling, Phenylmethyl-deactivated deactivated
5 Details
961-01482-04  SPME liner Base deact w/o packing
Special SPME liner, without filling, Base-deactivated deactivated
5 Details
961-01482-07  SPME liner Siltek deact w/o packing
Special SPME liner, without filling, Siltek-deactivated deactivated
5 Details

Injector Liner GC-2014 / GC-2025

Injector liner split injection for SPL-2014 and SPL-2025

Part number Description PCS Details
961-01485-02  SPL liner IPdeact. w.gl.wool
Standard split liner, prepacked with glass wool, Phenylmethyl-deactivated
5 Details
961-01485-03  SPL liner IPdeact. w.CarboFrit
Standard split liner, prepacked with CarboFrit, Phenylmethyl-deactivated
5 Details
961-01485-05  SPL liner Base deact. w.gl.wool
Standard split liner, prepacked with glass wool, Base-deactivated
5 Details
961-01485-06  SPL liner Base deact. w.CarboFrit
Standard split liner, prepacked with CarboFrit, Base-deactivated
5 Details
961-01485-08  SPL liner Siltek deact. w.gl.wool
Standard split liner, prepacked with glass wool, Siltek-deactivated
5 Details
961-01485-09  SPL liner Siltek deact. w. CarboFrit
Standard split liner, prepacked with CarboFrit, Siltek-deactivated
5 Details

Injector liner splitless injection for SPL-2014, WBI-2014, and SPL-2025

Part number Description PCS Details
961-01480-02  SPLless liner IPdeact. w.gl.wool
Standard splitless liner, prepacked with glass wool, Phenylmethyl-deactivated
5 Details
961-01480-03  SPLless liner IPdeact. w.CarboFrit
Standard splitless liner, prepacked with CarboFrit, Phenylmethyl-deactivated
5 Details
961-01480-05  SPLless liner Base deact. w.gl.wool
Standard splitless liner, prepacked with glass wool, Base-deactivated
5 Details
961-01480-06  SPLless liner Base deact. w.CarboFrit
Standard splitless liner, prepacked with CarboFrit, Base-deactivated
5 Details
961-01480-08  SPLless liner Siltek deact. w.gl.wool
Standard splitless liner, prepacked with glass wool, Siltek-deactivated
5 Details
961-01480-09  SPLless liner Siltek deact. w. CarboFrit
Standard splitless liner, prepacked with CarboFrit, Siltek-deactivated
5 Details

Further dimensions and specifications on request. For Research Use Only. Not for use in diagnostic procedures. This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.